Application of CLSI’s EP15A3 guideline in Accu-chek Inform II method verification

Authors

  • Hoang Bich Nga Le
  • Thi Thuy Quynh Tran, Lam Kieu Oanh Le
  • Thi Ngoc Lan Nguyen*

Keywords:

Accu-chek Inform II, blood glucose level, EP15A3, method validation.

Abstract

Method validation is an essential laboratory requirement to ensure reliable test results for diagnosis and treatment. This study aims to evaluate the determination of blood glucose level method by the Accu-chek Inform II analyzer at Hanoi Medical University Hospital. Method and materials: Internal control materials were provided by the manufacturer to conduct experiments to evaluate the precision, accuracy and linearity of blood glucose testing on the Accu-chek Inform II analyzer according to the guideline EP15A3 of the Clinical and Laboratory Standards Institute. Results: The within-run precision and within-laboratory precision of two experimental quality control levels are less than the manufacturer’s claim, the observed mean is not within the verification interval, but the Bias is less than the
allowable Bias. The accuracy and precision of the blood glucose test performed on the Accu-chek Inform II analyzer used at the Hanoi Medical University Hospital were confirmed in accordance with the manufacturer’s claims.

DOI:

https://doi.org/10.31276/VJST.65(7).43-46

Classification number

3.5

Author Biographies

Hoang Bich Nga Le

Hanoi Medical University, 1 Ton That Tung, Trung Tu Ward, Dong Da District, Hanoi, Vietnam

Hanoi Medical University Hospital, 1 Ton That Tung, Trung Tu Ward, Dong Da District, Hanoi, Vietnam

Thi Thuy Quynh Tran, Lam Kieu Oanh Le

Hanoi Medical University, 1 Ton That Tung, Trung Tu Ward, Dong Da District, Hanoi, Vietnam

Thi Ngoc Lan Nguyen*

Hanoi Medical University, 1 Ton That Tung, Trung Tu Ward, Dong Da District, Hanoi, Vietnam

Hanoi Medical University Hospital, 1 Ton That Tung, Trung Tu Ward, Dong Da District, Hanoi, Vietnam

Downloads

Published

2023-07-25

Received 5 October 2022, accepted 24 October 2022

How to Cite

Le Hoang Bich Nga, Tran Thi Thuy Quynh, Le Lam Kieu Oanh, & Nguyen Thi Ngoc Lan*. (2023). Application of CLSI’s EP15A3 guideline in Accu-chek Inform II method verification. Version B of Vietnam Journal of Science and Technology, 65(7). https://doi.org/10.31276/VJST.65(7).43-46

Issue

Section

Medical and Pharmacological Sciences